Password reset for Minimus₂/Minimus Lite

Scope

This procedure applies to Minimus₂ or Minimus Lite digitisers. It does not apply to Minimus digitisers or products containing them.

  • Minimus₂
    and Minimus Lite
This procedure only applies to these digitisers.
  • Minimus
This procedure does not apply to the Minimus digitiser, nor to Radian and Radian-string surface interface boxes. The passwords for these digitisers can be reset using the Minimus password reset procedure.
  • Minimus+
This procedure does not apply to the Minimus+ digitiser. The passwords for these digitisers can be reset using the Minimus password reset procedure.
  • Fortimus
This procedure does not apply to the Fortimus digitil accelerometer. The passwords for these digitisers can be reset using the Minimus password reset procedure.
  • Certimus
This procedure does not apply to the Certimus digital seismometer. The passwords for these digitisers can be reset using the Minimus password reset procedure.

Requirements

You will need:
  • Physical access to the unit to be upgraded.
  • A debug 'Y'-cable. This cable has two LEMO connectors and one DE9f connector.
Minimus debug cable
  • A PC or laptop with terminal emulator software. If you have Scream installed, you can use the built-in emulator; otherwise, we recommend PuTTY for windows systems and Minicom for Linux systems.
Generic laptop
  • If the PC or laptop does not have a 9-way COM port, a serial-to-USB converter. We recommend units based on the FTDI chip-set.
USB-to-serial converter

Expert Summary

ENsure that the firmware is at the latest revision and then use the debug 'Y'-cable to connect to the console at 115200, 8N1. Issue the following commands:

varuser set "Username (User)" user varuser set "Password (User)" user varuser set "Username (Admin)" admin varuser set "Password (Admin)" admin reboot

If in doubt, follow the detailed instructions below.

Detailed instructions

The following procedure stores the passwords for the user and admin accounts and then reboots the system. During the subsequent initialisation, the stored passwords will take effect.

Although this procedure applies to several different digitisers and instruments, the unit to be reset will be referred to as a Minimus₂ throughout.
  1. Start with the Minimus₂ powered up and running. You do not need network access to it.

  2. Disconnect the GNSS receiver from the Minimus₂ by grasping the sliding sheath and pulling in the direction of the cable.

    Unplugging the GNSS cable

    Do not twist and do not use tools.

    warning symbols
  3. Locate the male LEMO connector of the debug 'Y'-cable and plug it into the GNSS connector of the Minimus₂ by aligning the red dots and gently pushing the connector home until it clicks into place.

    Inserting a LEMO connector

  4. If required, connect the GNSS receiver to the other LEMO connector of the debug 'Y'-cable by aligning the red dots and gently pushing the connectors together until they click into place. This is not generally necessary: the Minimus₂ will not drift out of synchronisation if the GNSS receiver is unplugged for a few minutes.

    Connecting two in-line LEMOs

  5. Either

    Or

  6. Start your terminal emulation software in the usual way.

    Different emulators require different instructions. Use the radio-buttons below to select instructions which are specific to your terminal emulator:

    If you choose to use a different emulator, you will have to adapt the instructions below to suit your chosen software.

  7. Set the emulator to operate at 115,200 Baud, with eight data bits, no parity bits and one stop bit ('8N1').

  8. Enter the following five commands:

    varuser set "Username (User)" user varuser set "Password (User)" user varuser set "Username (Admin)" admin varuser set "Password (Admin)" admin reboot
  9. Exit the emulator:

  10. Disconnect the debug cable and replace the GNSS receiver.

Conclusion

This concludes the procedure. Once the Minimus₂ has finished rebooting, you should be able to access it using the default login details.

In case of any difficulty, please contact .